<?xml version="1.0" standalone="yes"?>
<?xml-stylesheet type="text/xsl" href="..\..\XSLTemplates\ErrorsStyleSheet.xsl"?>
<Error>
<ErrorNumber>16029</ErrorNumber>
<ErrorFamily>Image</ErrorFamily>
<ErrorMessage>
IMAGE: Failures detected when verifying SGE scan data
</ErrorMessage>
<AppliesTo>
<UnnumberedList>
<ListItem><Text>Green</Text></ListItem>
</UnnumberedList>
</AppliesTo>
<ServiceInformation>
<DescriptionText><text>
This error indicates a problem has been detected with the ECL memory on the SGE.  The ECL power
supplies (-5.2V and +4.6V) should be checked at the test points on the SGE.  If these are within spec,
reseat the SGE.  If the error persists, replace the SGE.
</text></DescriptionText>
</ServiceInformation>
<OperatorAction>
<DescriptionText><text>
Record error message and contact Creo Service
</text></DescriptionText>
</OperatorAction>
<Cause>
ECL power supplies out of spec
</Cause>
<Cause>
SGE improperly seated
</Cause>
<Cause>
SGE faulty
</Cause>
<SeeAlso>
<UnnumberedList>
</UnnumberedList>
</SeeAlso>
<OtherLinks>
<Link><LinkAddress>..\..\ShowAllErrors.htm</LinkAddress><LinkTitle>Firmware 0 Error Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllCommands.htm</LinkAddress><LinkTitle>Firmware 0 Command Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllNVS.htm</LinkAddress><LinkTitle>Firmware 0 NVS Index</LinkTitle></Link>
</OtherLinks>
<Updated>
Updated: Wed Jul 16 19:33:29 2008 PST
</Updated>
<Links>
</Links>
</Error>